Alarm Clear

10 Alarm clear line.

Alarms are cleared when a low level TTL signal is
applied.

A COM

11 This is the common line for the external signal pins

7 to 10, 12, 13, 19, 21, 24, and 25. It is connected
internally to the negative output.

PRL OUT+

12 Positive output line for master-slave parallel

operation.

Current Sum

13 Current signal line for master-slave parallel

operation.

Alarm Status

14 On when a protection function (OVP, HW OVP,

OCP, OHP, FAN, SEN, or AC_FAIL) has been
activated or when an output shutdown signal is
being applied (open-collector photocoupler
output).

1

PWR ON Status 15 Outputs a low level signal when power is turned

on. (open-collector photocoupler output).

1

OUT ON Status 16 On when the output is on (open-collector

photocoupler output).

1

TRIG OUT

17 Trigger signal output line.

N.C.

18 Not connected.

OUT ON/OFF
CONT

19 Output on/off line.

On when set to a low level TTL signal, Off when set
to a high level TTL signal. (F-94: 1)
On when set to a high level TTL signal, Off when
set to a low level TTL signal. (F-94: 0)

A COM

20 This is the common line for the external signal pins

7 to 10, 12, 13, 19, 21, 24, and 25. It is connected
internally to the negative output.