background image

PSU Series User Manual

58

4. Use the Current knob to set the F-

03 setting.

Set F-03 to 1 (CC High Speed
Priority) or 3 (CC Slew Rate
Priority) and save.

Current

F-03

1 = CC High Speed Priority
3 = CC Slew Rate Priority

5. Press the Voltage knob to save the

configuration setting. ConF will be
displayed when it is configuring.

Voltage

VOLTAGE

CURRENT

LAN

RMT

ERR

DLY

ALM

ISR

M 1

M 2

M 3

RUN

C V

C C

V

A

ISR indicator for CC Slew

Rate Priority (F-03=3)

6. If CC Slew Rate Priority was chosen as the

operating mode, set F-06 (Current Slew Rate
Up) and F-07 (Current Slew Rate Down) and
save.

F-06 / F-07 0.001A~2A / msec (PSU 6-200)

0.001A~1.2A / msec (PSU 12.5-120)
0.001A~0.76A / msec (PSU 20-76)
0.001A~0.38A / msec (PSU 40-38)
0.001A~0.25A / msec (PSU 60-25)

7. Press the Function key again to exit

the configuration settings. The
Function key light will turn off.

Function

8. Use the Voltage knob to set the

voltage limit (crossover point).

Voltage