CONFIGURATION

101

Falling Current
Slew Rate

Sets the falling current slew rate. Only
applicable if V-I Mode is set to CC Slew Rate
Priority.
F-07

0.001 ~ max. A/msec

Internal
Resistance
Settings

Sets the internal resistance of the power supply.
F-08

0.000Ω ~ X.XXXΩ

(Where X.XXX = Rating Voltage /
Rating Current)

Bleeder Control

Bleeder control turns ON/OFF the bleeder
resistor. Bleeder resistors discharge the filter
capacitors after power is turned off as a safety
measure.

F-09

0 = OFF, 1 = ON

Buzzer ON/OFF

Turns the buzzer sound on or off. The buzzer is
associated with alarm sounds and keypad entry
sounds.
F-10

0 = OFF, 1 = ON

OCP Delay Time

Sets the OCP delay time. This parameter will
delay the amount of time it takes to trigger the
over current protection. This function can be
useful to prevent current overshoot from
triggering OCP.
F-12

0.1 ~ 2.0 sec

Current Setting
Limit (I-limit)

Turns the current setting limit (I-limit) on or off.
Turning this function on will prevent you from
accidentally setting the current limit above the
set OCP level.
F-13

0 = OFF, 1 = ON

Voltage Setting
Limit

Turns the voltage setting limit (V-limit) on or
off. Turning this function on will prevent you
from accidentally setting the voltage limit
above the OVP level.