CONFIGURATION
101
Falling Current
Slew Rate
Sets the falling current slew rate. Only
applicable if V-I Mode is set to CC Slew Rate
Priority.
F-07
0.001 ~ max. A/msec
Internal
Resistance
Settings
Sets the internal resistance of the power supply.
F-08
0.000Ω ~ X.XXXΩ
(Where X.XXX = Rating Voltage /
Rating Current)
Bleeder Control
Bleeder control turns ON/OFF the bleeder
resistor. Bleeder resistors discharge the filter
capacitors after power is turned off as a safety
measure.
F-09
0 = OFF, 1 = ON
Buzzer ON/OFF
Turns the buzzer sound on or off. The buzzer is
associated with alarm sounds and keypad entry
sounds.
F-10
0 = OFF, 1 = ON
OCP Delay Time
Sets the OCP delay time. This parameter will
delay the amount of time it takes to trigger the
over current protection. This function can be
useful to prevent current overshoot from
triggering OCP.
F-12
0.1 ~ 2.0 sec
Current Setting
Limit (I-limit)
Turns the current setting limit (I-limit) on or off.
Turning this function on will prevent you from
accidentally setting the current limit above the
set OCP level.
F-13
0 = OFF, 1 = ON
Voltage Setting
Limit
Turns the voltage setting limit (V-limit) on or
off. Turning this function on will prevent you
from accidentally setting the voltage limit
above the OVP level.