Notes, Advanced mode (user-defined) – Teledyne LeCroy SierraFC M164 User Manual User Manual
Page 176
Teledyne LeCroy
Advanced Mode (User‐Defined)
176
SierraFC M164 Protocol Analyzer User Manual
Notes
Add a Project Note
To enter and save information about the current project, click the Notes tab and enter
the data about the project.
Figure 2.132: Project Notes Tab
Advanced Mode (User-Defined)
Advanced Mode expands Analysis capability by allowing you to program complex
triggering and data capture projects.
The Advanced Mode is a state machine with up to 23 different states. You can program
each state individually to:
Trigger on a different Event or trigger unconditionally.
Capture Everything, Nothing, or a user‐defined pattern.
Include up to three ELSE IF statements, allowing a jump to any other state based
on a user definition.
Use up to three timers, which you can set to a maximum value of 4294900 ms or
over one hour. If you enter a value larger than 42494900ms a warning pop‐up
displays: Invalid value! Please enter a value between 0 and 4294900.You can set
a timer in the state or continue the timer set in the previous state.
Output an external trigger High or Low.
Note:
In Advanced Mode, Events on each link are counted independently. A condition is met if the
number of Events on a link equals the defined occurrence.