beautypg.com

Figure 10-1. typical step response method, Table 10-4. step response method, Table 10-4 – Micromod Micro-DCI: 53SL5100A Single Loop Controller User Manual

Page 110: Cesses. see, Figure 10-1

background image

Single Loop Process Controller Instruction Manual

102 Instrument Tuning

Table 10-4. Step Response Method

Step

Procedure

1

Set the process to approximately normal conditions in Manual mode.

2

Introduce a 10% (approximate) step change to the manually regulated signal to the
valve. Observe the magnitude of the step in percent of full signal span F. Also plot a
graph of the resulting transient curve traced by the controlled variable.

3

Draw a straight line tangent to a point on the curve having the greatest slope (most
linear).

4

Measure the apparent dead time L in minutes as shown in

Figure 10-1

.

5

Measure RL, the product of the slope and dead time as shown in

Figure 10-1

. This

value should be expressed in percent of full scale span.

6

Compute the best settings as follows:
A. Proportional (P): P.B. = (100 RL)/ F.
B. Proportional & Integral(PI): P.B. = (100 RL)/ F, TR = 3L
C. Proportional, Integral, and Derivative (PID): P.B. = (80 RL)/ F, TR = 2L, TD = 0.5L

Figure 10-1. Typical Step Response Method