If a value outside of the range is entered, the
closest available value is automatically selected.

Nearest Available

Set vertical axis (Auto + Absolute)

Background

The Y (vertical) axis configuration is available for:

Manual or Automatic fit: Selects whether the
vertical range is manually set or automatically
adjusted as the graph is plotted.

Absolute or Percentage: Selects how the vertical
range is defined, as absolute values (minimum
and maximum) or percentage of the nominal
(center) value.

Panel operation

1. Press F2 to select Abs (Absolute).

Abs

%

F 2

2. Press F3 to select Auto Fit.

Auto Fit

F 3

3. Nothing new appears on the display: the LCR-

8000G automatically configures the vertical
scale according to the measured data.