GW Instek LCR-8101 (Replaced by LCR-8101G) User Manual
Page 29
GETTING
STARTED
29
10b. Select
Repetitive
measurement
Press the Sing/Rep key to select
Repetitive (automatic trigger)
measurement. Press the Left/Right arrow
key and move the cursor to Speed. Press
the Up/Down key to select the speed.
Page56
11a. Select
Absolute
measurement
Press F5 (Abs/%/∆) to select Abs. Press
the Left/Right key to move the cursor to
Lo (Low limit). Use the numerical and
unit keys to set the Low limit. Repeat this
for Hi (Hi limit) as well.
Page66
11b. Select
Percentage
measurement
Press F5 (Abs/%/∆) to select %. Press the
Left/Right key to move the cursor to the
Nominal value. Use the numerical and
unit keys to set the numerical level. Then
move the cursor to Lo (Low limit) and set
the percentage. Repeat this for Hi (Hi
limit) as well. To save the latest
measurement result as Nominal, press F6
(Save Nom).
Page67
11c. Select Delta
measurement
Press F5 (Abs/%/∆) to select ∆. Press the
Left/Right key to move the cursor to the
Nominal value. Use the numerical and
unit keys to set the numerical level. Then
move the cursor to Lo (Low limit) and set
the Lo value difference. Repeat this for Hi
(Hi limit) as well. To save the latest
measurement result as Nominal, press F6
(Save Nom).
Page68
Optional settings
To hide the drive Voltage/Current, press
the Code key, type 80, then press Enter.
Set the Range (internal setting) to Auto,
use the Left/Right key to move the cursor
and Up/Down key to change the setting.
Page57
Page51