Page 91

background image

GRAPH

MODE

91

Unit Mismatched

If a value outside of the range is entered, the
closest available value is automatically selected.

Nearest Available

If the entered start Voltage level is higher than
the stop Voltage, the two values are swapped.

Hi and Lo Swapped

Set stop Voltage

4. Repeat the above step for the stop Voltage.

Stop:

1.00 V

Range

10mV ~ 2V (AC≤3 MHz)
10mV ~ 1V (AC>3 MHz) *1mV step

(Stop Voltage must be higher than the Start
Voltage)

Set measurement
Frequency

5. Use the Up/Down key to move the

cursor to Frequency setting (Freq).

Freq :

10.000kHz

Use the numerical keys to enter the
measurement frequency.

Range

20Hz ~ 1MHz/5MHz/10MHz

50Hz

Enter

1MHz

M

Enter

Backspace

All clear

Clear

Select Lin/Log
scale

6. Press F1 to select the horizontal

scale, Linear or Logarithmic.

Log(V)

Lin(V)

F 1