LCR-8000G Series User Manual

62

Beep in repetitive
mode

If the repetitive measurement is On, the beep
might sound continuously. If this becomes a
problem, either use the Single mode (press
Sing/Rep key) or turn Off the beep.

Configure the Average

Background

The Average function sets the number of samples
used, which are then averaged as the final output.
The number of samples varies from 1 to 256.

Panel operation

1. Press the Menu key, then F5

(System). The system configuration
appears.

Menu

F 5

Precision LCR Meter LCR-8101

5

:

Frequency

Software version 2.03 Oct 25 2008

RS-232

GPIB

Graph mode

Line frequency

Beep

GPIB address

:

:

OFF

50Hz

1MHz

Average

: 10

2. Press the Up/Down key and move

the cursor to Average.

Average

:

10

3. Use the number pad to select

numerical keys to enter the number
of samples to be averaged. A
maximum of 256 samples can be
selected for averaging.