Pass/fail test (single step) – GW Instek LCR-8101 (Replaced by LCR-8101G) User Manual
Page 28
LCR-8000G Series User Manual
28
Pass/Fail test (Single step)
Step Description
Details
1. Connect fixture
Connect the fixture to the DUT.
Page24
2. Set buzzer
sound
Press the Menu key, then F5 (System).
Press the Up/Down arrow key to move
the cursor to Beep, then use the
Left/Right key to select setting (Off
recommended).
Page61
3. Set Average
Press the Menu key, then F5 (System).
Press the Up/Down arrow key to move
the cursor to Average, then use the
numerical keys to enter an average
number (1-256). Press enter to confirm the
number.
Page62
4. Enter menu
Press the Menu key, followed by F1 (AC
measurement) or F2 (Rdc).
Page47
5. Show scale
Press F4 (Show/Hide Scale) to show the
scale (or hide the circuit diagram)
Page 49
6. Select measur-
ement item
Press F1 (First) and F2 (Second)
repeatedly to select the measurement
item.
Page50
7. Select series/
parallel circuit
If available, press F3 (Series/Parallel) to
select the equivalent circuit model.
Page50
8. Set
measurement
frequency
Press the Left/Right arrow key and move
the cursor to the Frequency. Use the
numerical and unit keys to set the level.
Page52
9. Set
measurement
Voltage
Press the Left/Right arrow key and move
the cursor to the Voltage. Use the
numerical and unit keys to set the level.
Page54
10a. Select Single
measurement
Press the Sing/Rep key to select Single
(manual trigger) measurement. To trigger
measurement, press the Trig key.
Page55