Page 78

background image

LCR-8000G Series User Manual

78

Set Hi limit

Move the cursor to Hi column.
Enter the Hi limit using the
numerical keys and unit keys.

Spd
Hi

1.0000H

0.0000H

Lo

MAX

Range

follows the specification for each
measurement item

Example:

1.5kH

(for Ls)

Enter

H

k

Set Lo limit

Move the cursor to Lo column.
Enter the Lo limit using the
numerical keys and unit keys.

Spd
Hi

0.0000H

Lo

MAX

1.0000H

Range

follows the specification for each
measurement item

Example:

1.0kH

(for Ls)

Enter

H

k

Set trigger delay

Move the cursor to Dly (Delay)
column. Enter the amount of
trigger delay time using the
numerical keys and unit keys.

Hi
Lo

10 ms

Dly

1.0000H
0.0000H

Range

0 (no delay) ~ 1000ms

Example:

10ms

m