Calibration command – GW Instek LCR-8101 (Replaced by LCR-8101G) User Manual
Page 117

REMOTE
CONTROL
117
:multi:new
Create a new multi-step program.
Example: :multi:new demo (file name demo)
:multi:res?
Query the results of the test for each step.
Parameter: 0 (Pass), 1 (Fail Hi), 2 (Fail Lo)
Example: 1, +1.5E-7, 0, –0.2E-4 (step 1 failed on
high limit, step 2 passed)
:multi:run
Switches to the multi-step run page.
:multi:save
Save currently edited file.
:multi:speed
:multi:speed?
Sets or returns the measurement speed for the
currently selected step.
Send parameter: Max, Fast, Med, Slow
Return parameter: 0(Max), 1(Fast), 2(Med), 3(Slow)
Example: :multi:speed max (maximum speed)
:multi:test
:multi:test?
Selects or returns the step being edited.
Parameter: 1 ~ 30
Example: :multi:test 1(step 1 selected)
:multi:trig
Starts running multi-step measurements.
Calibration command
:cal:oc-trim
Performs open circuit trimming.
Parameter: 1 (Spot trim), 2 (<10kHz), 3 (<100kHz),
4 (All frequency)
Example: :cal:oc-trim 4 (calibrate for all frequency)
:cal:sc-trim
Performs short circuit trimming.
Parameter: 1 (Spot trim), 2 (<10kHz), 3 (<100kHz),
4 (All frequency), 5 (Rdc)
Example: :cal:sc-trim 4 (calibrate for all frequency)
:cal:res?
Returns the result of the calibration performed.
Parameter: 0 (fail), 1 (pass)