LCR-8000G Series User Manual

116

:multi:delay

:multi:delay?

Sets or returns trigger delay time for currently
selected step in millisecond.
Parameter: 0ms ~ 1000ms
Example: :multi:delay 10m (10ms)

:multi:freq

:multi:freq?

Sets or returns the frequency for the currently
selected step in Hz.
Parameter: 20 ~ 1/5/10MHz
Example: :multi:freq 1e3 (1kHz)

:multi:func

:multi:func?

Sets or returns measurement type for the currently
selected step.
Send parameter: LS, LP, Q, CS, CP, D, Z, PHASE,
RS, RP, X, G, B, Y, RDC
Return parameter: 1 (Z), 2 (Ls), 3 (Lp), 4 (Cs), 5
(Cp), 7 (Y), 8 (G), 9 (P), 10 (Q), 11 (D), 12 (Rs), 13
(Rp), 14 (B), 15 (X), 16 (Rdc)
Example: :multi:func ls (Series inductance)

:multi:hi-lim

:multi:hi-lim?

Sets or returns the higher test limit of the currently
selected step.
Example: :multi:hi-lim 10 (limit to 10.0)

:multi:lev

:multi:lev?

Sets or returns the drive level for the currently
selected step in Voltage.
Parameter: 10mV ~ 2V (DC/AC≤3 MHz)

10mV ~ 1V (AC>3 MHz)

Example: :multi:lev 200m (200mV)

:multi:load

Loads an existed file to run or edit.
Example: :multi:load demo (file name demo)

:multi:lo-lim

:multi:lo-lim?

Sets or returns the lower test limit of the currently
selected step.
Example: :multi:lo-lim –5 (limit to –5)