2. Press F1 (O/C Trim) or F2 (S/C

Trim) to select the trimming mode.

F 1

F 2

3. The trimming menu appears.

S/C

trimming

SHORT THE TEST LEADS

DC

Spot freq

F 1

F 2

F 6

Abort

SELECT A CHOICE
TO CONTINUE...

F 3

F 4

F 5

All freq

<= 100kHz

<= 10kHz

4. Always select F5 (All freq) when

using the standard fixture set
included in the LCR-8000G package.

F 5

DC

(S/C trimming only) Trims at 0Hz.

Spot

freq

Trims at the frequency set in the
measurement mode (page52).

<=

10kHz

Frequency range is 0Hz ~ 10kHz.

<=

100kHz

Frequency range is 0Hz ~ 100kHz.

All

freq

Frequency range is 0Hz ~ ≤1MHz.

Frequency limit example: When using a special
fixture set, trimming failed at 50kHz which is
outside of the test component range (up to
5kHz). In this case, use F3 (<= 10kHz).

5. The trimming starts and ends automatically.