APPENDIX

133

Specification

Test Frequency

LCR-8101G

DC, AC:20Hz~1MHz

LCR-8105G

DC, AC:20Hz~5MHz

LCR-8110G

DC,

AC:20Hz~10MHz

Basic Accuracy

R,Z,X,G,Y,B,L,C ±0.1%@1kHz

R&G Accuracy

When Qxؤ0.1, multiply Ae by √1+Q2 for R, G
accuracies.

Measurable
Parameters

Rac, Rdc, Rs, Rp, Z, Ls, Lp, D, G, B, θ, Cp, Cs, Q, Y, X

Measurement
Range

R,Z,X 0.1mΩ~100MΩ
G, Y, B

10ns~1ks

L

0.1nH~100kH

C

0.01pF~1F

D

0.00001~1000

Q

0.01~9999.9

Rdc

0.01mΩ~100MΩ

Ɵ

-180°~+180°

Equivalent
Circuit

Parallel

C+G, L+G, B+G, B+Q, B+D, B+R

Series

X+R, X+D, X+Q

Series & Parallel C+R, C+D, C+Q, L+R, L+D, L+Q

Polar Form

Z + Phase Angle, Y + Phase Angle

Input Impedance 100Ω
Measurement
Speed

DC

Max:30ms; Fast:60ms;
Medium:120ms; Slow:900ms

AC≤100Hz Max:600ms;

Fast:650ms;

Medium:1.2s; Slow:1.3s

AC≤2kHz

Max:120ms; Fast:180ms;
Medium:470ms; Slow:600ms

AC>2kHz Max:75ms;

Fast:150ms;

Medium:450ms; Slow:600ms

AC=1MHz
(LCR-8101G)

Max:120ms; Fast:150ms;
Medium:470ms; Slow:620ms

AC≥1~5MHz

(LCR-8105G)

Max:120ms; Fast:150ms;
Medium:470ms; Slow:620ms

AC≥1~10MHz
(LCR-8110G)

Max:120ms; Fast:150ms;
Medium:470ms; Slow:620ms