PASS-FAIL

MODE

77

Set frequency

Move the cursor to Freq column.
Enter the frequency using the
numerical keys and unit keys.

Func
Freq

500.00

Ls

2.00 V

Volt

Range

20Hz ~ 1MHz/5MHz/10MHz
5 digit resolution

Example:

0.5kHz (500Hz)

k

Enter

Set voltage

Move the cursor to Volt column.
Enter the voltage using the
numerical keys and unit keys.

Freq
Volt

2.00 V

500.00

Bias

Range

(1mV step)

10mV ~ 2V (DC or AC≤3 MHz)
10mV ~ 1V (AC>3 MHz)

Example:

100mV

m

V/A

Enter

Select data
capture rate

Move the cursor to Spd column.
Select the capture rate by
pressing F1 (Prog) repeatedly.

Bias
Spd

MAX

1.0000H

Hi

F 1

DC

AC≤
100Hz

AC≤
2kHz

AC>
2kHz

AC≥
1MHz

Slow

900ms 1.3s 600ms 600ms 620ms

Med

120ms 1.2s 470ms 450ms 470ms

Fast

60ms 650ms 180ms 150ms 150ms

Max

30ms 600ms 120ms 75ms 120ms