Chapter 26 error messages, 1 api error messages, Api error messages -1 – Yokogawa PC-Based MX100 User Manual
Page 1219
![background image](/manuals/360310/1219/background.png)
26-1
IM MX190-01E
Error Messages
26
Chapter 26 Error Messages
26.1 API Error Messages
When a function fails during execution, an error number is returned. Below is a list of
error numbers, message strings, and corrective actions. They are common to the
MX100 and DARWIN.
Error number Message string
Description
Corrective Action/Remarks
0
Success
Completed normally.
-
1
Communication error
Communication error occurred.
Check the communication environment (address, cable, power to the device).
2
Timeout
Communication timeout occurred.
Check the communication environment (load condition).
3
Receive continue
Receive data continues.
Received data too long. Receive the rest of the data or check the
communication procedure.
4
Creating connection is failure
Failed to create the communication descriptor.
Memory or resources may be low. Check the PC environment.
5
Creating descriptor is failure
Failed to create the device descriptor.
Memory or resources may be low. Check the PC environment.
6
Connection exists already
Communication already established.
Do not connect to a device that is already connected.
7
Not connected
Not connected.
May be executing a command without making a connection. Make a
connection, then execute the command.
8
Not descriptor
No device descriptor.
The device descriptor designation may be incorrect. Specify the device
descriptor.
9
Commands are not processed successfully
Failed to execute the command.
Error occurred on the measurement instrument. Check the transmitted
command or the operation mode of the main unit.