Auto calibration, 5 auto calibration (autocal, Calibration – Ohaus Voyager Balances User Manual

Page 88

background image

24

4.5 Auto calibration (AutoCal

TM

)

On balances equipped with the AutoCal

TM

feature, calibration can be accomplished using an internal calibration mass.

Auto calibration can be performed at any time providing the balance has warmed up to operating temperature.

CALIBRATE

SPAN

LINEARITY

USER

CALIBRATION TEST

AUTOCAL

AUTOCAL DELTA CORRECTION

EXIT TO WEIGH

From main menu

Scroll to CALIBRATION and press the Enter button. Display advances to
CALIBRATE with SPAN highlighted. Scroll to AUTOCAL.

Enter

INTERNAL CALIBRATION

PLEASE WAIT...

Please wait is displayed while the balance is calibrating. If the balance is
unstable, a message indicating this appears. The balance will continue self
calibration until successful.

INTERNAL CALIBRATION

PLEASE WAIT. . .

THE DIFFERENCE BETWEEN THIS CAL

AND THE LAST CAL IS: 0.000 GRAMS

PRESS TO CONTINUE

Display indicates if calibration was successful and the difference between
the last calibration. It is not necessary to press the Enter button, the bal-
ance returns to a weighing mode automatically.

0.000

BASIC WEIGHING

0% 100%

Balance returns to a weighing mode.

GRAMS STABLE

PRESS FOR MENU

INTERNAL CALIBRATION

PLEASE CLEAR THE PAN. . .

AND PRESS .

Clear the pan.

Enter

Calibration