Auto calibration, 5 auto calibration (autocal, Calibration – Ohaus Voyager Balances User Manual
Page 88
24
4.5 Auto calibration (AutoCal
TM
)
On balances equipped with the AutoCal
TM
feature, calibration can be accomplished using an internal calibration mass.
Auto calibration can be performed at any time providing the balance has warmed up to operating temperature.
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
AUTOCAL DELTA CORRECTION
EXIT TO WEIGH
From main menu
Scroll to CALIBRATION and press the Enter button. Display advances to
CALIBRATE with SPAN highlighted. Scroll to AUTOCAL.
Enter
INTERNAL CALIBRATION
PLEASE WAIT...
Please wait is displayed while the balance is calibrating. If the balance is
unstable, a message indicating this appears. The balance will continue self
calibration until successful.
INTERNAL CALIBRATION
PLEASE WAIT. . .
THE DIFFERENCE BETWEEN THIS CAL
AND THE LAST CAL IS: 0.000 GRAMS
PRESS
Display indicates if calibration was successful and the difference between
the last calibration. It is not necessary to press the Enter button, the bal-
ance returns to a weighing mode automatically.
0.000
BASIC WEIGHING
0% 100%
Balance returns to a weighing mode.
GRAMS STABLE
PRESS
INTERNAL CALIBRATION
PLEASE CLEAR THE PAN. . .
AND PRESS
Clear the pan.
Enter
Calibration