beautypg.com

16 printing data (cont.) span calibration printout, Calibration test printout, Linearity calibration printout – Ohaus Voyager Balances User Manual

Page 39

background image

33

3.16 Printing Data (Cont.)

Span Calibration Printout

When performing a Span calibration, a printout is auto-
matically made after the calibration mass is placed on
the pan and the Print button is pressed.

Calibration Test Printout

When performing a Calibration Test with GLP turned on,
a printout is available. When the display indicates the
mass value to be placed on the pan, the balance the
automatically displays the calibration weight required
and the Print button is pressed.

Linearity Calibration Printout

When performing a Linearity calibration with GLP turned
on, a printout is automatically made after the calibration
mass is placed on the pan and the Print button is
pressed.

- - - - - CAL TEST - - - - - -
7/01/96

1:00:00 PM

Bal Id 1234
Cal:

4000.00g

Act:

4000.04g

Dif:

0.04g

Wt. Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................

- - - - - END - - - - -

- - - - - SPAN CAL - - - - - -
7/01/96

1:00:00 PM

Bal Id 1234
Cal:

4000.00g

Old:

4000.00g

Dif:

0.00g

Wt. Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................

- - - - - END - - - - -

- - - - - LIN CAL - - - - - -
7/01/96

1:00:00 PM

Bal Id 1234
Cal:

4000.00g

Old:

3999.94g

Dif:

0.06g

Wt. Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................

- - - - - END - - - - -

SAMPLE PRINTOUTS