Installation

Wind Shield

On 610g to 6100g balances with 0.01g resolution, a wind shield is
required to reduce the possibility of air currents from disturbing the pan.
Make sure the wind shield is firmly snapped into place.

NOTE: 4100g, 6100g and 8100g balances with 0.1g resolution and
internal calibration are equipped with a 6” Pan and Wind Shield.

2.5 Wind shield

0.00

BASIC

WEIGHING

0%

100%

GRAMS

STABLE

Enter

Go

Back

Heip

Print

>O/T<

>O/T<

PRESS

FOR

MENU

2.6 Power supply

Check to ensure the voltage printed on the AC adapter matches your
local line voltage. If this is not the case, on no account connect the AC
adapter to the power supply, but contact your responsible OHAUS dealer.

Rear of Balance

AC Adapter Connection

Connect the AC Adapter supplied to the three pin connector located at the rear
of the balance.

The balance now performs a self-test, loads parameters, displays the
software version, capacity, indicates if autocal is installed and the normal
weight display then appears.

Ensure that the AC adapter can never come into contact with liquids!

Allow your balance to warm up for at least 30 minutes to enable it to
adapt itself to the ambient conditions. If the balance has been stored in a
very cold environment before installation, it may require several hours for
the balance to stabilize.