beautypg.com

Calibration, 2 span calibration, Procedure – Ohaus Voyager Balances User Manual

Page 14

background image

8

Go Back

Help

Print

Enter

Go Back

Help

Print

Enter

MAIN MENU

CHANGE UNITS

SET BALANCE

RETURN TO

WEIGH

CUSTOM MENU

CHANGE MODE

CALIBRATION

MAIN MENU

CHANGE UNITS

SET BALANCE

LIBRARY

CUSTOM MENU

CHANGE MODE

CALIBRATION

CALIBRATE

SPAN

LINEARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH

MAIN MENU

CHANGE UNITS

SET BALANCE

RETURN TO

WEIGH

CUSTOM MENU

CHANGE MODE

CALIBRATION

CALIBRATE

SPAN

LINPLEASEERARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH

INTERNAL CALIBRATION

PLEASE CLEAR THE PAN
AND PRESS < ENTER > . . .

MAIN MENU

CHANGE UNITS

SET BALANCE

LIBRARY

CUSTOM MENU

CHANGE MODE

CALIBRATION

CALIBRATE

SPAN

LINPLEASEERARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH

INTERNAL CALIBRATION

PLEASE WAIT . . .

MAIN MENU

CHANGE UNITS

SET BALANCE

LIBRARY

CUSTOM MENU

CHANGE MODE

CALIBRATION

CALIBRATE

SPAN

LINPLEASEERARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH

INTERNAL CALIBRATION

PLEASE WAIT . . .

MAIN MENU

CHANGE UNITS

SET BALANCE

LIBRARY

CUSTOM MENU

CHANGE MODE

CALIBRATION

CALIBRATE

SPAN
LINEARITY

USER

CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH

ENTER USER CAL VALUE

UP/DOWN TO SCROLL CHARACTERS
LEFT/RIGHT TO MOVE CURSOR
ENTER TO ACCEPT
BACK TO QUIT

0

SPAN CALIBRATION

SPAN CAL SUCCESSFULLY DONE!
THE LAST DIFFERENCE BETWEEN THIS CAL
AND LAST CAL IS:
0.01 GRAMS
PRESS ANY KEY TO EXIT . . .

PROCEDURE

• Press

arrow button to select MAIN MENU.

• Press Enter button.

• Using the arrow buttons, select CALIBRATION.

• Press Enter button.

• Press

or

arrow button and select SPAN.

• Press Enter button.

• Clear the pan and press Enter button.

• Place indicated mass value on pan and press Enter

button.

• Display indicates if calibration was successful and the

difference between the last calibration.

• Press any button, balance returns to weighing mode.

• Remove masses from the pan.

NOTE: The samples shown on the displays were for an
4.1kg balance.

3.3.2 Span Calibration

Span calibration normally requires that calibration be made using a mass equal to the full capacity of the balance,
however, the Voyager balance can be calibrated using other lesser values as specified on the display.

0.00

WEIGHT

0%

100%

GRAMS

STABLE

MOVE HIGHLIGHT

MAIN MENU

CONTRAST

MAIN MENU

CHANGE UNITS

SET BALANCE

LIBRARY

CUSTOM MENU

CHANGE MODE

CALIBRATION

CALIBRATE

SPAN

LINPLEASEERARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH

SPAN CALIBRATION

PLEASE PUT: 4000.00 G ON PAN.
AND PRESS < ENTER > . . .

OTHER WEIGHTS MAY BE USED:
3000 2000 1000

0.00

WEIGHT

0%

100%

GRAMS

STABLE

MOVE HIGHLIGHT

MAIN MENU

CONTRAST