beautypg.com

Specifications, 8 specifications analyticals, Precision top loaders – Ohaus Voyager Balances User Manual

Page 60

background image

54

Capacity (g)

62

110

210

100/210 *

Readability (mg)

0.1

0.1/1

Weighing modes

g, mg, oz , oz t, ct, dwt, taels (3), mommes, gn, ti, N, custom unit

Functions

Parts counting, differential weighing, quick check, statistics, formulation, filling, animal weighing, percent

formulation, check weighing, density determination, SQC & pipette calibration.

Performance enhancing features

Program-Link

TM,

GLP protocol, contrast and brightness control, custom menu, pop-up windows, application

library, go-back key, help text

Repeatability (Std. dev.) (mg)

0.1

0.1/0.5

Linearity (mg) (+)

0.2

0.2/0.5

Tare range

Full capacity by subtraction

Safe overload capacity

150% of capacity

Stabilization time

<4 seconds

Sensitivity drift PPM/°C

3

(10°C - 30°C)

Operating temperature range:
w/internal calibration

10° to 40°C/ 50° to 104°F

w/o internal calibration

10° to 30°C/50° to 86°F

Calibration

Auto calibration

Power requirements

External Adapter, 100 -120 V ac, 220 - 240 V ac, 50/60 Hz
Plug configuration for US, Euro, UK, Japan & Australia

Draft shield (in/cm)

10.2/25.9

(free height above platform)

Display (in/cm)

0.6/1.5

Pan size (in/cm)

3.5/9. diameter

Dimensions

9 x 15.25 x14/22.8 x 38.7 x 35.5

(WxHxD) (in/cm)

5.8 Specifications

Analyticals

Capacity (g)

210

410

100/410* 610 2100 4100 1000/4100* 4100 6100 8100

Readability (g)

0.001

0.001/0.01

0.01

0.01/0.1

0.1

Weighing modes

g, mg, kg, lbs, oz , oz t, ct, dwt, taels (3), mommes, gn, ti, N, custom unit

Functions

Parts counting, differential weighing, quick check, statistics, formulation, filling, animal weighing, percent

formulation, check weighing, density determination, SQC & pipette calibration.

Performance enhancing features

Program-Link

TM,

GLP protocol, contrast and brightness control, custom menu, pop-up windows, application

library, go-back key, help text

Repeatability (Std. dev.) (g)

0.0005

0.0005/0.005

0.005 0.01/0.05 0.05

Linearity (g) (+)

0.002

0.002/0.005

0.02 0.02/0.05 0.1

Tare range

Full capacity by subtraction

Stabilization time

<3 seconds

Sensitivity drift PPM/°C

4

3

4

3

4 3 4 3

(10°C - 30°C)

Operating temperature range:
w/internal calibration

10° to 40°C/50° to 104°F

w/o internal calibration

10° to 30°C/50° to 86°F

Calibration

Auto-calibration

Power requirements

External Adapter, 100 -120 V ac, 220 - 240 V ac, 50/60 Hz
Plug configuration for US, Euro, UK, Japan & Australia

Draft shield (in/cm)

10.2/25.9

(free height above platform)

Display (in/cm)

0.6/1.5

Pan size (in/cm)

4.7/12 Dia.

6.8 x 6.8/17.2 x 17.2 w/windshield

8 x 8/ 20.3 x 20.3

Dimensions

9 x 15.25 x14/22.8 x 38.7 x 35.5

8.25 x 4 x 14/23.4 x 10.1 x 35.5

(WxHxD) (in/cm)

Precision Top Loaders

NOTE: Not all weighing modes apply depending upon capacity and resolution of the balance.

* Moveable FineRange

TM