Rs232 parameters, Setting the readout, 3 setting the readout – Ohaus Voyager Balances User Manual
Page 304: Setting averaging level, Setting stability level, Setting auto-zero, Setting legal for trade
63
Balance settings
7.3 Setting the readout
This menu option enables you to set the balance averaging level, stability level, auto zero, and legal for trade
(LFT).
Setting averaging level
Press Enter button, AVERAGING LEVEL FILTER is displayed. Averaging level
filter compensates for vibration or excessive air currents. A zero setting
(GOOD) corresponds to minimum filtering with fastest stabilization time. A
setting of (BEST)10 is maximum filtering with the slowest stabilization time.
Adjust to desired setting and press Enter button.
Setting stability level
Scroll to STABILITY LEVEL and press Enter button, STABILITY LEVEL
FILTER is displayed. Stability level filter determines the variation range in
divisions for a given reading depending upon the filter setting. A setting of
zero (GOOD) permits the balance to respond to variations of .5 divisions. A
setting of (BEST) 10 sets the threshold to 3 divisions. Adjust to desired
setting and press Enter button.
READOUT SETUP
AVERAGING LEVEL FILTER
GOOD
BEST
5
or
READOUT SETUP
STABILITY LEVEL FILTER
GOOD
BEST
5
or
Setting auto-zero
Scroll to AUTO ZERO and press Enter button, AZT is displayed. Auto zero
minimizes the effects of temperature changes and shift on the zero reading.
There are 4 settings, OFF, 0.5, 1, and 3. The numbers represent display
divisions. The balance maintains the zero display until the set threshold is
exceeded. Select the desired setting and press Enter button.
READOUT SETUP
AZT
or
OFF
0.5
1
3
Setting legal for trade
Scroll to LEGAL FOR TRADE and press Enter button. Legal for trade can be
set ON or OFF. When set ON, only the functions allowed by the national
weights and measures legislation are available. Press Enter button after
making your selection.
READOUT SETUP
AVERAGING LEVEL 5
STABILITY LEVEL 5
AUTO ZERO 0.5
LEGAL FOR TRADE OFF
EXIT TO WEIGH
or
SET BALANCE
Enter
From main menu
Select SET BALANCE and press the Enter button. Display advances to SET
BALANCE with READOUT highlighted. Press the Enter button, display
advances to READOUT SETUP.
READOUT