Balance modes

Select RUN. Place animal on the pan, press Enter button, START SAMPLE is
highlighted at the bottom of the display. An averaged reading is displayed
after a count-down. The count-down time is affected by the averaging level
filter setting.

6.3 Animal weighing

This feature permits weighing small animals directly on the balance. If an animal cage is used, the balance can be
tared (weight of cage is subtracted) and then the subject is placed in the cage and weighed. An averaging filter
compensates for animal movements. The filter can be set in a range from Good to Best. The large numeric display
indicates the weight of the subject and the bar graph indicates balance capacity used.

MAIN MENU

CHANGE MODE

ANIMAL WEIGHING SETUP

Enter

From main menu

Enter CHANGE MODE. Select ANIMAL WEIGHING, display advances to
ANIMAL WEIGHING SETUP.

ANIMAL WEIGHING SETUP

AVERAGING LEVEL 5

AUTO SAMPLE

OFF

RUN

Enter

Enter AVERAGING LEVEL, display advances to AVERAGING LEVEL FILTER.

AVERAGING LEVEL FILTER

GOOD

BEST

5

or

A setting of GOOD is used for subjects which are inactive. The balance will
respond quickly and present an average reading. When a setting of BEST is
used, the balance responds slowly providing a much longer averaging time
for very active subjects. Select filtering level and press Enter button.

ANIMAL WEIGHING SETUP

AVERAGING LEVEL 5

AUTO SAMPLE

OFF

RUN

Enter

AUTO SAMPLE when selected and set ON permits placing subjects on pan
one after another with the weight showing automatically after each subject.
When OFF is selected, user must press Enter button after each subject is
placed on the pan.

ANIMAL WEIGHING SETUP

AVERAGING LEVEL 5

AUTO SAMPLE

OFF

RUN

Enter

When AUTO SAMPLE has been selected, it is not necessary to press the
Enter button to activate sampling.