Calibration, Span calibration, 1 span calibration – Ohaus Voyager Balances User Manual

Page 84

background image

20

4.1

Span calibration

Span calibration normally requires that calibration be made using a mass equal to the full capacity of the

balance, however, the Voyager balance can be calibrated using other lesser values as specified on the

display.

CALIBRATE

SPAN

LINEARITY

USER

CALIBRATION TEST

AUTOCAL

AUTOCAL DELTA CORRECTION

EXIT TO WEIGH

From main menu

Scroll to CALIBRATION and press the Enter button. Display advances to
CALIBRATE with SPAN highlighted.

Enter

SPAN CALIBRATION

PLEASE CLEAR THE PAN. . .

AND PRESS .

Clear the pan.

Enter

SPAN CALIBRATION

PLEASE PUT 400.000 GRAMS

ON PAN AND PRESS

OTHER WEIGHTS MAY BE USED:

300 200 100

Enter

Place indicated mass value on pan.

SPAN CALIBRATION

PLEASE WAIT. . .

THE DIFFERENCE BETWEEN THIS CAL

AND THE LAST CAL IS: 0.000 GRAMS

PRESS TO CONTINUE

Enter

Display indicates the difference between the present calibration and the last
calibration.

400.000

GRAMS STABLE

BASIC WEIGHING

Enter

PRESS FOR MENU

0% 100%

Display indicates weight of the calibration masses. Remove masses from
pan. Balance returns to a weighing mode.

Calibration