beautypg.com

3 linearity calibration, Procedure – Ohaus Voyager Balances User Manual

Page 15

background image

9

Go Back

Help

Print

Enter

• Press

arrow button to select MAIN MENU.

• Press Enter button.

• Using the arrow buttons, select CALIBRATION.

• Press Enter button.

• Press

or

arrow button and select LINEAR-

ITY.

• Press Enter button, LINEARITY CALIBRATION is dis-

played.

• Clear pan and press Enter button.

• Place indicated mass value on pan. This will be one

half of the total capacity of the balance, then press
Enter button.

• Place indicated mass value on pan. This will be the

total capacity of the balance, then press Enter button.

PLEASE WAIT is displayed followed by the actual weight
value placed on the pan which is the maximum capacity
of the balance.

• Remove the masses from the pan. The balance is

now calibrated.

NOTE: The samples shown on the displays were for an
4.1kg balance.

PROCEDURE

Go Back

Help

Print

Enter

MAIN MENU

CHANGE UNITS

SET BALANCE

RETURN TO

WEIGH

CUSTOM MENU

CHANGE MODE

CALIBRATION

MAIN MENU

CHANGE UNITS

SET BALANCE

RETURN TO

WEIGH

CUSTOM MENU

CHANGE MODE

CALIBRATION

CALIBRATE

SPAN

LINEARITY

USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH

LINEARITY CALIBRATION

PLEASE CLEAR THE PAN
AND PRESS ENTER KEY . . .

MAIN MENU

CHANGE UNITS

SET BALANCE

RETURN TO

WEIGH

CUSTOM MENU

CHANGE MODE

CALIBRATION

CALIBRATE

SPAN

LINEARITY

USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH

LINEARITY CALIBRATION

PLEASE WAIT . . .

MAIN MENU

CHANGE UNITS

SET BALANCE

LIBRARY

CUSTOM MENU

CHANGE MODE

CALIBRATION

CALIBRATE

SPAN

LINEARITY

USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH

LINEARITY CALIBRATION

PLEASE PUT 2000.00 G ON PAN
AND PRESS < ENTER > . . .

MAIN MENU

CHANGE UNITS

SET BALANCE

RETURN TO

WEIGH

CUSTOM MENU

CHANGE MODE

CALIBRATION

CALIBRATE

SPAN

LINEARITY

USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH

LINEARITY CALIBRATION

PLEASE WAIT . . .

MAIN MENU

CHANGE UNITS

SET BALANCE

LIBRARY

CUSTOM MENU

CHANGE MODE

CALIBRATION

CALIBRATE

SPAN

LINEARITY

USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH

LINEARITY CALIBRATION

PLEASE PUT 4000.00 G ON PAN
AND PRESS < ENTER > . . .

3.3.3 Linearity Calibration

Linearity calibration utilizes three calibration points, one at zero, center span and full span. This method minimizes
deviation between actual and displayed weights within the balance's weighing range. Three weight values are used;
zero, a weight value at midpoint of the balance's weighing range and a weight value at or near the specified capacity.

0.00

WEIGHT

0%

100%

GRAMS

STABLE

MOVE HIGHLIGHT

MAIN MENU

CONTRAST

4000.00

WEIGHT

0%

100%

GRAMS

STABLE

MOVE HIGHLIGHT

MAIN MENU

CONTRAST

MAIN MENU

CHANGE UNITS

SET BALANCE

LIBRARY

CUSTOM MENU

CHANGE MODE

CALIBRATION

CALIBRATE

SPAN

LINEARITY

USER
CALIBRATION TEST
AUTOCAL