beautypg.com

Rs232 parameters, Setting the interface, 4 setting the interface – Ohaus Voyager Balances User Manual

Page 128: Setting baud rate, Setting data bits, Setting parity, Setting stop bits

background image

64

7.4 Setting the interface

This menu option enables you to set the balance communication parameters for the RS232 interface which include:
baud rate, data bits, parity and stop bit.

Setting baud rate

Baud rates of 300, 1200, 2400, 4800 and 9600 are available for commu-
nications. Select the appropriate rate. Default setting is 2400 baud. Press
Enter button.

INTERFACE

BAUD RATE

or

300

1200

2400

4800

9600

RETURN

Setting data bits

Data bits of 7 or 8 are selectable. Select appropriate bit rate. Default setting
is 7. Press Enter button.

INTERFACE

DATA BITS

or

7

8

RETURN

Setting parity

Select appropriate parity setting of either NONE, EVEN, ODD, 0, or 1. Default
setting is NONE. Press Enter button.

INTERFACE

PARITY

or

NONE

EVEN

ODD

0

1

RETURN

Setting stop bits

Select appropriate stop bits setting of either 1 or 2. Default setting is 2.
Press Enter button.

INTERFACE

STOP BITS

or

1

2

RETURN

SET BALANCE

Enter

From main menu

Select SET BALANCE and press the Enter button. Display advances to SET
BALANCE with READOUT highlighted. Scroll to INTERFACE and press the
Enter button, display advances to INTERFACE.

INTERACE

Balance settings