ENTER LIQUID DENSITY IN CC

Enter density of test liquid

Enter the density value for the test liquid. The standard density for water has
been pre-entered. Enter the correct value for the liquid you are using, and
press Enter button. Display advances to RETURN TO SETUP.

0.9971

0.9971

Return to setup

Press the Enter button. Display advances to PIPETTE SETUP with BARO
PRESSURE highlighted.

SELECT TEST LIQUID

H2O

OTHER

RETURN TO SETUP

Enter

Enter

PIPETTE SETUP

LIBRARY NAME

P2

TEST LIQUID(G/CC)

0.9971

BARO PRESSURE

1.0ATM

AUTO SAMPLE

OFF

AUTO PRINT

OFF

NUMBER OF SAMPLES

10

PIPETTE NAME

Enter barometric pressure

Press the Enter button. The display advances to BAROMETRIC PRESSURE
with ATM highlighted.

Select unit of atmospheric pressure measurement

Select unit, press Enter button. The display advances to BAROMETRIC
PRESSURE.

BAROMETRIC PRES-

SURE

ATM
PSIG
MMHG
MBAR
HPA
INHG

Select auto sampling

If you want auto sampling, press Enter button. Display advances AUTO
SAMPLE with ON highlighted. Press the Enter button.

Enter

BAROMETRIC PRESSURE

Specify barometric pressure

Enter the barometric pressure of the test area, then press the Enter button.
Display advances to PIPETTE SETUP with AUTO SAMPLE OFF highlighted. If
you do not want auto sampling, skip next two steps and select AUTO PRINT.

1.0000

1.000

It should be noted that in the next step you can select AUTO SAMPLE either
ON or OFF. When AUTO SAMPLE is turned ON, it is a real time saver as the
balance controls do not have to be touched during sampling.

PIPETTE SETUP

LIBRARY NAME

P2

TEST LIQUID(G/CC)

0.9971

BARO PRESSURE

1.0ATM

AUTO SAMPLE

OFF

AUTO PRINT

OFF

NUMBER OF SAMPLES

10

PIPETTE NAME

Select auto print on

Press the Enter button. The display advances to PIPETTE SETUP with AUTO
PRINT OFF highlighted.

Enter

PIPETTE SETUP

LIBRARY NAME

P2

TEST LIQUID(G/CC)

0.9971

BARO PRESSURE

1.0ATM

AUTO SAMPLE

OFF

AUTO PRINT

OFF

NUMBER OF SAMPLES

10

PIPETTE NAME

Balance modes