86

A-2 Porous material density determinations

The density of a porous (oil impregnated part) can be made with the balance. Weigh the part (dry) prior to oil
impregnation and record its weight. You must also know the density value of the oil to be used in immersing the
part before starting. In this procedure, you will follow the method for solid density measurements using water.

From main menu

Enter CHANGE MODE. Select DENSITY, display advances to LIBRARY NAME.

MAIN MENU

CHANGE MODE

Enter

DENSITY SETUP

LIBRARY NAME

Library name

Enter a library name, up to 8 characters and press Enter button. Display
advances to DENSITY SETUP with SOLID DENSITY highlighted. If a previous
name was entered and selected, the balance uses the parameters which
were previously set.

DENSITY SETUP

Enter

LIBRARY NAME

D2

SOLID DENSITY

<-SELECTED

LIQUID DENSITY

AUTO SAMPLE

OFF

AUTO PRINT

OFF

RUN

Select SOLID DENSITY, display advances to AUX LIQ & MODE with H2O
highlighted.

AUX LIQ & MODE

H20

<-SELECTED

OTHER

POROUS MATERIAL

OFF

DRY WEIGHT

O.000

DENSITY OF OIL

0.0000

RETURN TO SETUP

Enter

Selecting water as auxiliary liquid

Select H2O, display advances to AUX LIQ & MODE with ENTER TEMPERA-
TURE CELSIUS highlighted.

AUX LIQ & MODE

ENTER TEMPERATURE CELSIUS

Specify temperature

Enter the temperature of the water and press Enter button. Display
advances to AUX LIQ & MODE with POROUS MATERIAL highlighted.

AUX LIQ & MODE

Specify porous material

Select Porous Material and press Enter button. Display advances to POROUS
MATERIAL ON/OFF. Select ON and press Enter button.

H20

<-SELECTED

OTHER

POROUS MATERIAL OFF

DRY WEIGHT

O.000

DENSITY OF OIL

0.0000

RETURN TO SETUP

Enter

AUX LIQ & MODE

H20

<-SELECTED

OTHER

POROUS MATERIAL ON

DRY WEIGHT

200.040

DENSITY OF OIL

0.0000

RETURN TO SETUP

Enter dry weight

Enter the dry weight of the part as previously measured. Press Enter button.
Display advances to DENSITY OF OIL highlighted.