4.3 User calibration
User calibration is used when it is desired to calibrate the balance using a mass of known value. To use this calibra-
tion feature, proceed as follows:
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
AUTOCAL DELTA CORRECTION
EXIT TO WEIGH
From main menu
Scroll to CALIBRATION and press the Enter button. Display advances to
CALIBRATE with SPAN highlighted. Scroll to USER.
Enter
USER CALIBRATION
USER CALIBRATION RANGE:
100-400 GRAMS
PRESS User calibration range is specified. Enter USER CALIBRATION PLEASE PUT 200.000 GRAMS ON PAN AND PRESS Enter Place selected mass value on pan. USER CALIBRATION PLEASE WAIT. . . THE DIFFERENCE BETWEEN THIS CAL AND THE LAST CAL IS: 0.000 GRAMS PRESS Enter Display indicates the difference between the present calibration and the last 200.000 BASIC WEIGHING Enter 0% 100% Display indicates weight of the calibration masses. Remove masses from USER CALIBRATION ENTER CAL VALUE IN GRAMS 200.000 200.000 Enter the calibration value. GRAMS STABLE PRESS USER CALIBRATION PLEASE CLEAR THE PAN. . . AND PRESS Clear the pan. Enter Calibration
calibration.
pan. Balance returns to a weighing mode.