beautypg.com

2 advanced counting, Function link, Library name – Ohaus Voyager Balances User Manual

Page 273: Apw/size, Tare weight, Auto optimization

background image

32

Balance modes

6.1.2 Advanced counting

Advanced Count contains a number of entries which include assigning a library name, filling and sorting applications
and statistical information which can be printed. Once a library name is assigned, this particular counting function
can be recalled at any time.

PARTS COUNTING

ADVANCED PARTS COUNTING

PART COUNTING SETUP

LIBRARY NAME

Enter

Function link

Four options are available, OFF, FILLING, CHECK WEIGHING and STATISTICS.

When FILLING is selected, a target weight is entered which is shown as

100% on the bar graph on the display. When material is added to the

balance pan, it is displayed as a percentage and weight. When CHECK

WEIGHING is selected, a separate display has entries for nominal pieces,

over pieces, under pieces, run and exit. This type of function permits check-

ing of individual pieces against the stored information in the balance. When

STATISTICS is selected, provides display of Standard Deviation, either popu-

lation or sample, Mean, Sum, High, Low and Difference readings. Each can

be individually set ON or OFF. RUN when selected starts the program.

Library name

Enter library name, up to 8 characters and press Enter button. The library
name can be recalled later and the same type of items can be counted at
any time. The display advances to APW/SIZE.

APW/SIZE

APW/Size

Enter APW/SIZE. You have a choice of entering a sample size or average
piece weight. Enter required data. Display advances to TARE WEIGHT.

PART COUNTING SETUP

TARE WEIGHT

Tare weight

Enter the tare weight. This is the tare weight of the container holding the
samples.

PARTS COUNTING SETUP

or

AUTO OPTIMIZATION ON

Auto optimization

An ON or OFF function. When set ON, optimizes the accuracy based on
sample size. After selecting, the display advances to FUNCTION LINK.

FUNCTION LINK

OFF

FILLING

CHECK WEIGHING

STATISTICS

PARTS COUNTING SETUP
APW/SIZE

TARE WT

AUTO OPTIMIZATION

FUNCTION LINK

RUN

From main menu

Enter CHANGE MODE. Select PARTS COUNTING and then select ADVANCED
PARTS COUNTING. Display advances to PARTS COUNTING SETUP with
LIBRARY NAME highlighted.

Enter

SAMPLE SIZE

SET A.P.W.