4.2 Linearity calibration
Linearity calibration utilizes three calibration points, one at zero, center span and full span. This method minimizes
deviation between actual and displayed weights within the balance's weighing range. Three weight values are used;
zero, a weight value at midpoint of the balance's weighing range and a weight value at or near the specified capac-
ity.
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
AUTOCAL DELTA CORRECTION
EXIT TO WEIGH
From main menu
Scroll to CALIBRATION and press the Enter button. Display advances to
CALIBRATE with SPAN highlighted. Scroll to LINEARITY.
Enter
LINEARITY CALIBRATION
PLEASE CLEAR THE PAN. . .
AND PRESS
Clear the pan.
Enter
LINEARITY CALIBRATION
PLEASE PUT 200.000 GRAMS
ON PAN AND PRESS Enter Place indicated mass value on pan. LINEARITY CALIBRATION PLEASE WAIT. . . THE DIFFERENCE BETWEEN THIS CAL AND THE LAST CAL IS: 0.000 GRAMS PRESS Enter Display indicates the difference between the present calibration and the last 400.000 GRAMS STABLE BASIC WEIGHING Enter PRESS 0% 100% Display indicates weight of the calibration masses. Remove masses from LINEARITY CALIBRATION PLEASE PUT 400.000 GRAMS ON PAN AND PRESS Enter Place indicated mass value on pan. Calibration
calibration.
pan. Balance returns to a weighing mode.