Calibration
4. Calibration
Your balance can be calibrated with an internal mass (optional) or external masses. The balance can also be
checked by a test with internal or external masses. If you have a printer attached to your balance, the data of the
calibration and the results are printed out following GLP recommendations. A choice of six calibration methods are
available: Span, Linearity, User, Test, Automatic AutoCal
TM
and AutoCal
TM
with Delta Correction.
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
AUTOCAL DELTA CORRECTION
EXIT TO WEIGH
Span -
Span calibration ensures that the balance reads correctly
within specifications using two weight values: zero and a
weight value at 100% of the balances full capacity.
Linearity -
Linearity calibration minimizes deviation between actual
and displayed weights within the balances weighing
range. Three weight values are used: zero, a weight
value at midpoint of the balances weighing range, and a
weight value at or near the balances specified capacity.
User-
User calibration is a method where the balance can be
calibrated using a mass of known value by entering that
value into the balance.
Test-
Calibration test allows the stored calibration data to be
tested against the current mass being used for the test.
AutoCal
TM
On balances equipped with AutoCal
TM,
automatic calibra-
tion of the balance is accomplished by an internal mass.
AutoCal
TM
On balances equipped AutoCal
TM
, a software feature
W/Delta
allows the internal calibration mass to be adjusted ± 100
Correction
divisions at full scale capacity. This permits adjusting
the balance using an external Class I mass which is
traceable to a certified standard.