57

CHANGE MODE

PIPETTE SETUP

LIBRARY NAME

Enter

Library name

Enter a Library Name, up to 8 characters and press Enter button. Display
advances to PIPETTE SETUP with TEST LIQUID G/CC highlighted. If a previ-
ous name was entered and selected, the balance uses the parameters which
were previously set.

6.11 Pipette calibration

Pipette calibration checks the accuracy and precision values of pipettes by weight analysis. An analytical balance is
recommended for maximum accuracy. The balance is capable of recording data from 3 to 30 samples of each
pipette tested. Each test run is stored in the application library. The number of tests which can be stored will depend
on the number of samples per test. The density table for water is included. If other liquids are used for pipette
calibration, you must enter the liquid’s density in g/cc at current room temperature. Since all calculations are made
within the balance, it is also required that you know the atmospheric pressure which has to be entered. A printout
can be made which specifies all parameters of the test made.

From main menu

Enter CHANGE MODE and select PIPETTE, display advances to LIBRARY
NAME.

PIPETTE

Enter

PIPETTE SETUP

LIBRARY NAME

P2

TEST LIQUID(G/CC)

0.9971

BARO PRESSURE

1.0ATM

AUTO SAMPLE

OFF

AUTO PRINT

OFF

NUMBER OF SAMPLES

10

PIPETTE NAME

Test liquid

Press Enter button. Display advances to SELECT TEST LIQUID with H2O
highlighted.

Selecting water as the test liquid

If you are using water as the test liquid, press the Enter button. If you are
using a liquid other than water, skip the next step and continue.

SELECT TEST LIQUID

H2O

OTHER

RETURN TO SETUP

Enter

ENTER TEMPERATURE CELSIUS

Enter temperature of water

Measure the temperature of water you will be using for Pipette measure-
ments in Celsius. A standard room temperature of 25 degrees Celsius has
been pre-entered in the balance. Enter the correct value.

25.000

25.000

Selecting other test liquid

If you are using a liquid other than water as the test liquid, scroll to OTHER
and press the Enter button. The display advances to ENTER LIQUID DENSITY
IN CC.

SELECT TEST LIQUID

H2O

OTHER

RETURN TO SETUP

Balance modes