Configuration menu/selecting setup, C2 calibration sub-menu – Hardy HI 3010 Filler/Dispenser Controller User Manual
Page 76

HI-3010 Filler/Dispenser/IBC
60
Service Manual
Step 21. A “function OK” momentarily appears on the
screen indicating the calibration was successful
•
A message that says “Function Error” means
that the calibration was not successful.
Check Chapter 7 - Troubleshooting of this
manual for corrective action.
•
Another message may occur which is: Secu-
rity Violation. This means that the User does
not have the security level required to do a
calibration.
Step 22. Press the Exit button until you return to the Standby
display.
Step 23. C2 calibration is complete.
C2 Calibration From the Web Page
Step 1. On the Filler/Dispenser Home Page Click on Con-
figuration. (See Fig. 5-11) The Configuration page
appears. (See Fig. 5-12)
FIG. 5-11 CONFIGURATION MENU/SELECTING
SETUP
FIG. 5-12 CONFIGURATION PAGE
Step 2. Click on Calibration. The Calibration Sub-menu
appears. (See Fig. 5-13)
FIG. 5-13 C2 CALIBRATION SUB-MENU
Step 3. The Load Sensor number is a read only field. It tells
you how many load sensors are connected to the
instrument. If this number does not reflect with the
actual number of load sensors go to Chapter 7 -
Troubleshooting.
Step 4. To choose the type of sensor for this instrument
click on the pull down list next to Sensor Type. (See
Fig. 5-14)