beautypg.com

Unit of measure parameters, About unit of measure, Decimal point parameter – Hardy HI 3010 Filler/Dispenser Controller User Manual

Page 54: About the decimal point parameter, Totalizer menu/choosing ingredient, Totalizer number menu, Clearing all totals for all 12 ingredients, Setup menu/unit of measure

background image

HI-3010 Filler/Dispenser/IBC

38

Service Manual

Step 4. To clear the Totalizer for all 12 ingredients or for

each individual ingredient press the Enter button.
The Totalizer Menu appears. (See Fig. 4-71)

FIG. 4-71 TOTALIZER MENU/CHOOSING

INGREDIENT TOTALIZER NUMBER

Step 5. Press the left or right arrow buttons to select the

Ingredient Totalizer you want to view.

Step 6. Press the Enter button. The Totalizer Number Menu

appears with the cursor in front Total Wt. (See Fig.
4-72) In our example we selected Ingredient Total-
izer #1.

FIG. 4-72 TOTALIZER NUMBER MENU

Step 7. Total WT. and Total CT. are read only.
Step 8. To Clear the Totals for Ingredient Number 1:

Press the down arrow until the cursor is in
front of Clear Totalizer.

Press the Enter button. The Total WT and
Total CT entries turn to 00.00.

Step 9. Press the Exit button to return to the Totalizer

Menu.

Step 10. To clear all totals for all 12 ingredients, press the

down arrow until the cursor is in front of Clear All
Totals. (See Fig. 4-73)

FIG. 4-73 CLEARING ALL TOTALS FOR ALL 12

INGREDIENTS

Step 11. Press the Enter button. All totals are cleared.
Step 12. Press the Exit button to return to the SETUP

MENU.

Unit of Measure Parameters

About Unit of Measure

The Unit of Measure Parameter sets the scale to either Eng-
lish or Metric units. The Selections are:

Pounds (lb)

Ounces (oz)

Kilograms (kg)

Grams (g)

NOTE:

Changing the units of measure converts all
parameter settings to the selected units.

PARAMETER:UNIT OF MEASURE
RANGE: LB, KG. OZ, G
DEFAULT: LB

Step 1. Press the down arrow button until the cursor is in

front of Unit of Measure. (See Fig. 4-74)

FIG. 4-74 SETUP MENU/UNIT OF MEASURE

Step 2. Press the right or left arrow buttons to make your

selection.

Step 3. Press the Enter button to set the entry.

Decimal Point Parameter

About the Decimal Point Parameter

The Decimal Point Parameter is set to determine the resolu-
tion you want for a fill/dispense. Here you set the location of
the decimal point for the weight resolution. The higher the
number the farther to the left the decimal point moves and
the higher the resolution of the scale. It is important to note
that setting the resolution does effect the overall accuracy of
the instrument. Increasing the number of decimal points
increases the overall accuracy of the instrument.

PARAMETER:DECIMAL POINT
RANGE: 0-6
DEFAULT: 2