Hardy HI 3010 Filler/Dispenser Controller User Manual
Page 118

HI-3010 Filler/Dispenser/IBC
102
Service Manual
C1 - Guidelines for Instabilities on formerly operating systems with Smart Diagnostics
TEST
COMPLETE
Stable?
Yes
No
C1
Repeat for each sensor.
Record load sensor for
comparison and stability
Check individual load sensors output by
using the Smart Diagnostics and
SD summing card
A load sensor output can be considered stable if
the readings only vary +- 0.01mV. Applying
weight and releasing the weight should show a
very rapid Millivolt change. A slow Millivolt
change can indicate a damaged strain guage.
Inspect the summing junction box for
contamination or damage, and replace if necessary
Stable?
Stable?
Insure the problem is not
mechanical. (Review Section B)
If you are unable to isolate the
problem, contact Hardy Technical
Support: 800-821-5831
Yes
No
No
No
Insure the C2 wires are correctly terminated
even if you do not have C2 load sensors.
The IT communications and testing is
conducted over the C2 lines.
Yes
Remove and replace any load
sensor determined to be unstable