FIG. 4-83 SETUP MENU/TARE LIMIT

Step 2. Press the Clear button to clear the current entry.
Step 3. Use the alphanumeric key pad to enter the new Tare

Limit value.

Step 4. Press the Enter button to set the entry.

Averages Parameter

About the Averages Parameter

This setting is to aid in ignoring the effects of material
impact. If material is not entering or exiting the scale evenly,
weight fluctuations can be seen. Applications requiring very
quick weight readings should reduce this setting to it’s mini-
mum. If the weight is unstable due to material impacting,
increase the averages. This sets the number of weight read-
ings that will be used to compute the displayed weight. The
average is a sliding average so that a new average is avail-
able for display at every reading.

The Filler/Dispenser does 55 updates per second which
translates to an update approximately every 20 milliseconds.
If you average enough weight readings the weight loss or
gain remains smooth. If you average the weight too much
you can cause over filling. Here is an example of 5 averages
reading 5 Engineering Units (EU):

PARAMETER:AVERAGES
RANGE: 1-250
DEFAULT: 1

Step 1. Press the Down arrow until the cursor is in front of

Averages. (See Fig. 4-84)

FIG. 4-84 SETUP MENU/AVERAGES

Step 2. Press the right or left arrow buttons to increase or

decrease the number of averages.

Step 3. Press the Enter button to set the entry.

Scale Capacity Parameter

About the Scale Capacity Parameter

If this value is exceeded by 5% a HI indication appears on
the front display. Communications to and from optional
devices are not effected. This value is the nominal operating
capacity of the scale. (It is recommended that you use the
default parameter)

PARAMETER:SCALE CAP
RANGE: .000001-999999
DEFAULT: 999999

Step 1. Press the Down arrow button until the cursor is in

front of Scale Cap. (See Fig. 4-85)

FIG. 4-85 SETUP MENU/SCALE CAPACITY

Step 2. Press the Clear button to clear the current entry.
Step 3. Use the alphanumeric key pad to enter the new

Scale Capacity value.

Step 4. Press the Enter button to set the entry.

The WAVERSAVER

®

Parameter

About the WAVERSAVER Parameter

Typically, mechanical noise (from other machinery in a plant
environment) is present in forces larger than the weight
forces trying to be detected. The Filler/Dispenser is fitted

with WAVERSAVER

®

technology which eliminates the

20ms

20ms

20ms

20ms

20ms

0

0

0

0

5

0

0

0

5

5

0

0

5

5

5

0

5

5

5

5

5

5

5

5

5

1

2

3

4

5 = 5 EU in 100ms

TABLE 4-1: 5 AVERAGES READING 5 EU