beautypg.com

Scale – Delta MMC120 User Manual

Page 110

background image

110

SCALE

Default: 30300

Range: -32767 to 32767, excluding 0

TIP: From the Tools menu, select Scale/Offset Calibration to help calculate

SCALE and

OFFSET

.

The SCALE parameter is used in calculating the

ACTUAL POSITION

in Position Units

from the

TRANSDUCER COUNTS

, using the following formula (Position Units can be

0.001", 0.1 mm, etc.):

ACTUAL POSITION =

TRANSDUCER COUNTS x SCALE

+

OFFSET

32768 x Prescale Divisor

In linear position applications, SCALE accounts for the following:

1.Transducer Compensation: The SCALE parameter compensates for differences in

magnetostrictive transducers. Each transducer will indicate a different physical

distance for a given interrogation cycle. Usually, only a small change from the

default value is necessary to compensate the transducer.

2. Distance Translation: A second and more useful feature of SCALE is its ability to

translate a fixed physical distance to usable position units (position units from

thousandths to hundredths or vice versa). The translation is done using a combination

of SCALE and recirculation selection.

SCALE is defined as 32768 times the number of Position Units per

TRANSDUCER

COUNT

times the Prescale Divisor, as shown in the following equation (P0 and P1 are

two positions, C0 and C1 are the corresponding

TRANSDUCER COUNTS

, and the

Prescale Divisor is specified in the

CONFIGURATION WORD

):

SCALE =

(P0 – P1)

x 32768 x Prescale Divisor

(C0 - C1)

The SCALE can be calculated on a magnetostrictive displacement transducer using its

calibration number (nominally 9.012 microseconds per inch) with the following formula:

SCALE =

Position Units per Inch x 32768 x Prescale Divisor

x Sign

Cal. Number (

µ

s/in) x 120MHz x # of

Recirculations

In the above formula, the 120MHz comes from the motion module’s internal counter, and

Sign

equals +1 when an extend move yields increasing

ACTUAL POSITION

counts,

and -1 when an extend move yields decreasing

ACTUAL POSITION

counts.

The calibration number must be specified in the same units as the Position Units (inches,

mm, etc.). Position Units are generally 0.001 inches or 0.1 mm. Usually there are 1000

Position Units per inch when the calibration number is in microseconds per inch.