Valves

Instrument Guide

| 55

Column Switching Valve

The NGC system supports up to three column switching valves. Each column

switching valve is used to connect up to five columns simultaneously, supporting up

to 15 columns. This enables automated column scouting, and the use of different

columns without connecting and disconnecting the columns. The valve includes an

internal bypass mode, which enables buffers to bypass the connected columns

when necessary (for example, when priming or cleaning the system). The valve can

also reverse the flow, which is optimal for certain separations where elution in a

narrow band is necessary. Reverse flow is also useful when cleaning a column.

The column switching valve has one inlet port, one outlet port, and five pairs of

column connection ports. In ChromLab you can program the method to

automatically change the position of the valve. The active ports are identified by

LEDs.

The column switching valve has two internal pressure sensors. ChromLab displays

the precolumn and delta pressure values. The pressure values are viewed as traces

on the chromatogram. When two or three column switching valves are present, the

NGC system monitors the delta pressure from all columns that are inline.

Overpressure events are triggered when any of the delta pressures reaches its

maximum specified limit.