For information about changing the uv lamps, see, Replacing the multi – Bio-Rad ChromLab™ Software User Manual
Page 211
The Detectors
Instrument Guide
| 209
Replacing the Multi-Wavelength UV/Vis Detector
Lamps
This section explains how to replace the deuterium lamp or tungsten lamp inside the
multi-wavelength UV/Vis detector. The replacement lamp unit includes the lamp,
cable connector, and plug. The lamp is replaced as a unit.
Caution:
It is strongly suggested that you wear gloves when you replace the
lamps. Do not touch the lamp glass with bare hands, as oils from your skin will
degrade the lamp over time.
To replace the multi-wavelength UV/Vis lamps
1.
On the touch screen, select Shut Down on the dropdown menu to exit
ChromLab and shut down the NGC instrument.
2.
Disconnect the tubing lines from the top and bottom of the UV cell.
3.
Disconnect the tubing lines from the conductivity monitor.
4.
Remove the multi-wavelength UV/Vis detector module from the instrument.
Replacing or Repositioning Modules on the NGC Instruments on page
for information about removing a module.
5.
Using a screwdriver, loosen the screw and remove the door on the left side of
the module.