Reassembling the mixer – Bio-Rad ChromLab™ Software User Manual
Page 196

A
| Maintaining the Instrument
194
| NGC Chromatography Systems and ChromLab Software
Reassembling the Mixer
Tip:
To reinsert the magnetic stir bar, place it on a tabletop and carefully place
the mixer base over it. Ensure that the magnetic stir bar lies flat in the mixer
base.
To reassemble the mixer
1.
Place the mixer top onto the top of the mixer base assembly.
Note:
Ensure that the four bolt holes on the bottom of the mixer top align
with the four bolt holes on the top of the mixer base assembly.
2.
Insert the four hex bolts into the bolt holes on the bottom of the mixer base.