Bio-Rad ChromLab™ Software User Manual
Page 259

UV or UV/Vis Detector
Instrument Guide
| 257
Sample trace values
are lower than
expected.
The lamps might be nearing the
end of their usable life.
Check the lamp usage values in
the System Information dialog
box. From the System Control
tab in ChromLab open File >
System Information > Detectors
tab.
If the lamp usage values are
close to their limits
(approximately 2,000 hours for
the deuterium and tungsten
lamps; approximately 5,000
hours for LEDs), change the
lamps.
For single-wavelength LEDs,
see
Wavelength UV Detector LED
on page
For multi-wavelength lamps,
see
Wavelength UV/Vis Detector
Lamps on page
Symptom
Possible Cause
Possible Solution