Bio-Rad ChromLab™ Software User Manual
Page 110

3
| Preparing the Instrument
108
| NGC Chromatography Systems and ChromLab Software
3.
Loosen all three screws on the top of the connector.
Tip:
These screws adjust the terminals that hold the wires in place. The
screws do not dislodge from the connector.
4.
Locate the analog output cable that ships with your detector.
5.
If connectors are attached to both ends of the cable, remove the connector that
does not insert into your detector.
Tip:
Alternatively, contact the manufacturer for a suitable cable.
6.
If necessary, strip the plastic wire coating off the wires so that they are
exposed.
7.
Obtain a small piece of 20–24 AWG negative wire.
8.
Bend the wire into a U shape and strip the plastic coating off both ends to
create a jumper wire.