beautypg.com

Disassembling the mixer – Bio-Rad ChromLab™ Software User Manual

Page 190

background image

A

| Maintaining the Instrument

188

| NGC Chromatography Systems and ChromLab Software

Disassembling the Mixer

Use this procedure to replace any of the components of the mixer.

Important:

Before disassembling the mixer, flush any hazardous material from

the NGC system. Ensure that the pumps are not running and bleed residual

pressure from the system.

To disassemble the mixer

1.

On the touch screen, select Shut Down on the dropdown menu to exit
ChromLab and shut down the NGC instrument.

2.

Disconnect the tubing from the top, bottom, and sides of the mixer.