beautypg.com

Bio-Rad ChromLab™ Software User Manual

Page 202

background image

A

| Maintaining the Instrument

200

| NGC Chromatography Systems and ChromLab Software

9.

Tighten the bolts. If necessary, use the screwdriver.

10. Reconnect the 1/4-28 fittings from the tubing to the top and bottom of the flow

cell and ensure that they are secure.

11. Reconnect the tubing lines to the conductivity monitor.

12. Restart the NGC instrument.