| Maintaining the Instrument
238
| NGC Chromatography Systems and ChromLab Software
Repositioning the Touch Screen
This procedure explains how to move the touch screen from one side of the NGC
instrument to the other.
To reposition the touch screen
1.
On the touch screen, select Shut Down on the dropdown menu to exit
ChromLab and shut down the NGC instrument.
2.
Disconnect the monitor cable from the touch screen connector port on the
back of the instrument.
Unscrew the thumbscrews and firmly pull the cable out of the connector port.
3.
Locate the Phillips screwdriver that is provided in the accessories kit.