Bio-Rad ChromLab™ Software User Manual
Page 155
Recommended Maintenance Schedule
Instrument Guide
| 153
Yearly
Sample pump seals and
check valve
Replace the seals and check valves using
the pump head maintenance kit.
System pump seals and
check valve
Replace the seals and check valve using the
pump head maintenance kit.
Post-inject valve tubing
Replace all tubing.
Inline filter
Replace if installed.
Bottle filter
Replace or clean if installed.
pH electrode
Replace yearly or earlier if necessary.
Check the values for slope and offset. Start
ChromLab software and select Tools >
Calibrate and choose pH from the dropdown
list. Complete the calibration steps in the
dialog box. Replace the pH probe if
necessary.
Store the pH electrode in storage solution
when not in use. See Table
for
recommended storage solutions.
Tip:
See
the electrode.