Sample loop psi ratings for u.s. and canada, Electromagnetic compatibility (emc) – Bio-Rad ChromLab™ Software User Manual
Page 300
E
| Regulatory Information
298
| NGC Chromatography Systems and ChromLab Software
Sample Loop PSI Ratings for U.S. and Canada
The following pressure ratings are required specifically for U.S. and Canada
applications. Bio-Rad recommends using these ratings to reduce the
volume-pressure product of the column load liquid path in order to maintain
compliance with UL61010-1:2012 and CSA C22.2 No. 61010-1-12.
Electromagnetic Compatibility (EMC)
CE Mark:
EN61326 Class A Electrical Equipment for Measurement, Control, and
Laboratory Use, General Requirements
The NGC instrument is designed to operate in a controlled electromagnetic
environment. Do not use radio transmitters such as mobile telephones within 10 ft
(3
m) of the instrument.
Table 24.
Sample loop PSI ratings for U.S. and Canada applications
Injection Loop Size
Pump Size
Maximum Pressure
Rating
5 ml
F10
2,650 psi
5 ml
F100
1,400 psi
DynaLoop™ 25 ml
F100
700 psi
DynaLoop 90 ml
F100
270 psi