Purging the ngc scout system – Bio-Rad ChromLab™ Software User Manual
Page 126

3
| Preparing the Instrument
124
| NGC Chromatography Systems and ChromLab Software
Purging the NGC Scout System
To purge the buffer blending valve
1.
Open the System Pump dialog box and select Priming in the Mode dropdown
list.
To purge an individual line of air bubbles, proceed to step 2 that follows.
To purge all lines of air bubbles sequentially, proceed to step
2.
To purge an individual line of air bubbles and fill it with buffer:
a.
Select a port to open.
b.
(Optional) Change the default flow rate.
The default flow rate is set to 20 ml/min.
c.
Click Start.
d.